Beckman Coulter Inc.: DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520
Recall · Food and Drug Administration
In FDA’s enforcement report of Oct. 14, 2026. Class II; Ongoing.
The record
- Recall number
- Z-0084-2027
- Agency
- Food and Drug Administration
- Firm
- Beckman Coulter Inc.
- Product
- DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520
- Hazard
- Beckman Coulter has identified a software issue affecting DxC 500 AU Clinical Analyzers when testing plasma samples. Following user configuration of a Clinical Chemistry test to use units different from the system default, a subsequent configuration change and activation of the draft configuration in the System Configuration page may cause the analyzer to revert the test units to the system default setting without notifying the user. The system does Pre-Review Reminders Include OII and CDRHRecallGroup@fda.hhs.gov in any discussions on change/elevation of recalls to Class I classification and situations involving additional public communications (FDA and/or firm issued). Post-Review Reminders Final Check: Ensure your documentation accurately reflects the final recommendation prior to signature. Complete CTS: Recall Product tab, Additional Recall Event Info, Regulation, and Act. Upload PDFs of Consult, Review Memo, and Classification Letter for Class I Recall in DocMan. [Click to enter extra reminder]To add reminders, type the reminder in the text field to the left then press the Tab key. Doc ID 03534.05.06 2 of 8 not provide a warning or notification to alert the user that the configured units have reverted to the default units. This issue can result in falsely elevated reported results.
- Units
- 146 US, 170 OUS
- Classification
- Class II
- Status
- Ongoing
- Distribution
- Worldwide - US Nationwide distribution.
- Who started it
- Voluntary: Firm initiated
The document
The Food and Drug Administration’s record of the recall is its data, not a document: its record here is the data as read.
What its source changed
Its source has changed nothing since Dorworth News first read it, .
The record as data
The record as data: JSON, 3.2 KB: Beckman Coulter Inc.: DxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520
The recall’s record: the product, the hazard, the remedy and the firm in the agency’s own words, its classification and status, and every change the agency made to it.
Source: openFDA (Food and Drug Administration), read . Data provided by the U.S. Food and Drug Administration.