  "id": "dorworth:recall:fda-z-0165-2027",
  "kind": "recall",
  "schemaVersion": 1,
  "schema": "https://dorworth.com/schemas/v1/recall.json",
  "url": "https://dorworth.com/relationships-family/recalls/fda-z-0165-2027",
  "title": "Howmedica Osteonics Corp.: TS Femoral Prep Tray, Triathlon. Catalog Number: 6543-6-507",
  "place": [
    {
      "scope": "nation",
      "code": "US",
      "name": "United States"
    }
  ],
  "source": {
    "body": "Food and Drug Administration",
    "url": "https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0165-2027%22",
    "retrievedAt": "2026-11-24T19:45:38.310-05:00"
  },
  "keptCopy": null,
  "published": "2026-11-24T19:45:38.310-05:00",
  "updated": "2026-11-24T19:45:38.310-05:00",
  "changes": {
    "count": 0,
    "url": "https://dorworth.com/changes?record=dorworth%3Arecall%3Afda-z-0165-2027"
  },
  "terms": {
    "credit": "Dorworth News",
    "url": "https://dorworth.com/terms"
  },
  "body": {
    "identifier": "Z-0165-2027",
    "issuedOn": "2026-10-21",
    "issuedBy": "Food and Drug Administration",
    "status": "listed",
    "sourcePage": "https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0165-2027%22",
    "sourceFile": null,
    "copyNote": "The Food and Drug Administration’s record of the recall is its data, not a document: its record here is the data as read.",
    "history": [],
    "agency": "fda",
    "category": "other",
    "firm": "Howmedica Osteonics Corp.",
    "product": "TS Femoral Prep Tray, Triathlon. Catalog Number: 6543-6-507.",
    "hazard": "Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.",
    "remedy": null,
    "units": "582 trays",
    "classification": "Class II",
    "recallStatus": "Ongoing",
    "notes": [
      {
        "text": "These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.",
        "label": "Distribution"
      },
      {
        "text": "Voluntary: Firm initiated",
        "label": "Who started it"
      }
    ]
  }
}