Medline Industries, LP: Medline surgical convenience kits labeled as: 1) VEIN HARVESTING PACK-LF, Medline product number/SKU DYNJ0826732J; 2) NBMC VASCULAR …
Recall · Food and Drug Administration
In FDA’s enforcement report of Oct. 21, 2026. Class II; Ongoing.
The record
- Recall number
- Z-0171-2027
- Agency
- Food and Drug Administration
- Firm
- Medline Industries, LP
- Product
- Medline surgical convenience kits labeled as: 1) VEIN HARVESTING PACK-LF, Medline product number/SKU DYNJ0826732J; 2) NBMC VASCULAR PACK-LF, Medline product number/SKU DYNJ0665837G; 3) AV FISTULA, Medline product number/SKU DYNJ900454C; 4) VEIN PACK SHD, Medline product number/SKU DYNJ68420D; 5) PACK VASCULAR BYPASS, Medline product number/SKU DYNJ905692J; 6) PRODUCT VASCULAR LAFAYETTE, Medline product number/SKU DYNJ908620D; 7) CHRISTUS SPOHN AV ACCESS PACK, Medline product number/SKU DYNJ60973A; 8) TOL VEIN PACK, Medline product number/SKU DYNJ80861F; 9) THROMBECTOMY PACK-LF, Medline product number/SKU DYNJ0160348B; 10) VEIN PACK, Medline product number/SKU DYNJ67308A; 11) PACK VASCULAR BYPASS, Medline product number/SKU DYNJ905692I; 12) PACK VASCULAR BYPASS, Medline product number/SKU DYNJ905692G; 13) BYPASS ACCESSORY, Medline product number/SKU DYNJ52036F; 14) TOL VEIN PACK, Medline product number/SKU DYNJ80861C; 15) MANSFIELD ACCESS GRAFT, Medline product number/SKU DYNJ909985; 16) VEIN HARVESTING PACK-LF, Medline product number/SKU DYNJ0826732I; 17) CAROTID ENDARTERECTOMY PK, Medline product number/SKU DYNJ85516; 18) RADIAL ARTERY PACK, Medline product number/SKU DYNJ83938; 19) AV FISTULA PACK, Medline product number/SKU DYNJ01966N; 20) FH A V FISTULA PACK-LF, Medline product number/SKU DYNJ51804G; 21) VEIN PACK SHD, Medline product number/SKU DYNJ68420C; 22) VEIN PACK SHD, Medline product number/SKU DYNJ68420B; 23) AV FISTULA, Medline product number/SKU DYNJ906945; 24) VEIN PACK SHD, Medline product number/SKU DYNJ68420A; 25) MHC AV FISTULA, Medline product number/SKU DYNJ903705G.
- Hazard
- Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
- Units
- 8374 kits
- Classification
- Class II
- Status
- Ongoing
- Distribution
- US Nationwide distribution.
- Who started it
- Voluntary: Firm initiated
The document
The Food and Drug Administration’s record of the recall is its data, not a document: its record here is the data as read.
What its source changed
Its source has changed nothing since Dorworth News first read it, .
The record as data
The record as data: JSON, 3.6 KB: Medline Industries, LP: Medline surgical convenience kits labeled as: 1) VEIN HARVESTING PACK-LF, Medline product number/SKU DYNJ0826732J; 2) NBMC VASCULAR …
The recall’s record: the product, the hazard, the remedy and the firm in the agency’s own words, its classification and status, and every change the agency made to it.
Source: openFDA (Food and Drug Administration), read . Data provided by the U.S. Food and Drug Administration.