Medline Industries, LP: Medline surgical convenience kits labeled as: 1) SPINAL CDS, Medline Product Number/SKU CDS940087AM; 2) CUSTOM CRANI, Medline Product …
Recall · Food and Drug Administration
In FDA’s enforcement report of Oct. 21, 2026. Class II; Ongoing.
The record
- Recall number
- Z-0179-2027
- Agency
- Food and Drug Administration
- Firm
- Medline Industries, LP
- Product
- Medline surgical convenience kits labeled as: 1) SPINAL CDS, Medline Product Number/SKU CDS940087AM; 2) CUSTOM CRANI, Medline Product Number/SKU DYNJ907697C; 3) FEM POP PACK-LF, Medline Product Number/SKU DYNJ0159820F; 4) AV FISTULA PACK, Medline Product Number/SKU DYNJ67139D; 5) CRANIOTOMY CDS, Medline Product Number/SKU CDS780025AD; 6) LAMINECTOMY PACK, Medline Product Number/SKU DYNJ17333K; 7) EXTREMITY-LF, Medline Product Number/SKU DYNJ905276F; 8) CRANIOTOMY CDS, Medline Product Number/SKU CDS780025AC; 9) SPINAL CDS, Medline Product Number/SKU CDS940087AL; 10) CUSTOM CRANI, Medline Product Number/SKU DYNJ907697A; 11) NEURO-SPINE BASIN PACK, Medline Product Number/SKU DYNJ80055; 12) CRANIOTOMY-RICHMOND, Medline Product Number/SKU CDS980929U; 13) ACB EXTRAS PACK-6211, Medline Product Number/SKU DYNJ49516; 14) SPINAL CDS, Medline Product Number/SKU CDS940087AK; 15) SPINAL CDS, Medline Product Number/SKU CDS940087AJ; 16) CRANIOTOMY CDS, Medline Product Number/SKU CDS780025AB; 17) SPINAL CDS, Medline Product Number/SKU CDS940087AI; 18) LAMINECTOMY PACK, Medline Product Number/SKU DYNJ17333J; 19) SPINAL CDS, Medline Product Number/SKU CDS940087AG; 20) CRANIOTOMY CDS, Medline Product Number/SKU CDS780025AA.
- Hazard
- Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
- Units
- 6069 kits
- Classification
- Class II
- Status
- Ongoing
- Distribution
- US Nationwide distribution.
- Who started it
- Voluntary: Firm initiated
The document
The Food and Drug Administration’s record of the recall is its data, not a document: its record here is the data as read.
What its source changed
Its source has changed nothing since Dorworth News first read it, .
The record as data
The record as data: JSON, 3.3 KB: Medline Industries, LP: Medline surgical convenience kits labeled as: 1) SPINAL CDS, Medline Product Number/SKU CDS940087AM; 2) CUSTOM CRANI, Medline Product …
The recall’s record: the product, the hazard, the remedy and the firm in the agency’s own words, its classification and status, and every change the agency made to it.
Source: openFDA (Food and Drug Administration), read . Data provided by the U.S. Food and Drug Administration.