Medtronic MiniMed, Inc.: CareLink Clinic, REF: MMT-7350
Recall · Food and Drug Administration
In FDA’s enforcement report of Oct. 28, 2026. Class II; Ongoing.
The record
- Recall number
- Z-0203-2027
- Agency
- Food and Drug Administration
- Firm
- Medtronic MiniMed, Inc.
- Product
- CareLink Clinic, REF: MMT-7350
- Hazard
- CareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which the calculation logic fails to use custom values entered by healthcare professionals (HCPs) for Total Daily Dose (TDD) or Long-acting insulin when calculating the Carb Ratio, Insulin Sensitivity Factor (ISF), and Max Recommended value for Rapid-acting insulin. Instead, the calculator continues to use the system-recommended values, resulting in incorrect rapid-acting insulin parameters. If an HCP or patient inputs these incorrect setting values into the patient's MiniMed Go app, the patient could deliver higher or lower insulin doses compared to their insulin needs, potentially leading to over/under delivery of insulin with a severity ranging from Catastrophic (death from severe hypoglycemia) to Minor (self-treatable hypoglycemia/hyperglycemia).
- Units
- 1
- Classification
- Class II
- Status
- Ongoing
- Distribution
- US Nationwide distribution in the state of KS.
- Who started it
- Voluntary: Firm initiated
The document
The Food and Drug Administration’s record of the recall is its data, not a document: its record here is the data as read.
What its source changed
Its source has changed nothing since Dorworth News first read it, .
The record as data
The record as data: JSON, 2.7 KB: Medtronic MiniMed, Inc.: CareLink Clinic, REF: MMT-7350
The recall’s record: the product, the hazard, the remedy and the firm in the agency’s own words, its classification and status, and every change the agency made to it.
Source: openFDA (Food and Drug Administration), read . Data provided by the U.S. Food and Drug Administration.