  "id": "dorworth:recall:fda-z-0207-2027",
  "kind": "recall",
  "schemaVersion": 1,
  "schema": "https://dorworth.com/schemas/v1/recall.json",
  "url": "https://dorworth.com/relationships-family/recalls/fda-z-0207-2027",
  "title": "TECHNO-PATH MANUFACTURING LTD.: Techno-Path Multichem AE Unassayed Level 3, REF:AE623X",
  "place": [
    {
      "scope": "nation",
      "code": "US",
      "name": "United States"
    }
  ],
  "source": {
    "body": "Food and Drug Administration",
    "url": "https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0207-2027%22",
    "retrievedAt": "2026-11-24T19:45:38.310-05:00"
  },
  "keptCopy": null,
  "published": "2026-11-24T19:45:38.310-05:00",
  "updated": "2026-11-24T19:45:38.310-05:00",
  "changes": {
    "count": 0,
    "url": "https://dorworth.com/changes?record=dorworth%3Arecall%3Afda-z-0207-2027"
  },
  "terms": {
    "credit": "Dorworth News",
    "url": "https://dorworth.com/terms"
  },
  "body": {
    "identifier": "Z-0207-2027",
    "issuedOn": "2026-10-28",
    "issuedBy": "Food and Drug Administration",
    "status": "listed",
    "sourcePage": "https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0207-2027%22",
    "sourceFile": null,
    "copyNote": "The Food and Drug Administration’s record of the recall is its data, not a document: its record here is the data as read.",
    "history": [],
    "agency": "fda",
    "category": "other",
    "firm": "TECHNO-PATH MANUFACTURING LTD.",
    "product": "Techno-Path Multichem AE Unassayed Level 3, REF:AE623X",
    "hazard": "A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.",
    "remedy": null,
    "units": "832 units",
    "classification": "Class II",
    "recallStatus": "Ongoing",
    "notes": [
      {
        "text": "US Nationwide distribution in the state of Massachusetts.",
        "label": "Distribution"
      },
      {
        "text": "Voluntary: Firm initiated",
        "label": "Who started it"
      }
    ]
  }
}