Abiomed, Inc.: Automated Impella Controller, Packaged. Primary user control interface for the Impella Catheter, catheter-based heart pump
Recall · Food and Drug Administration
In FDA’s enforcement report of Nov. 4, 2026. Class I; Ongoing.
The record
- Recall number
- Z-0208-2027
- Agency
- Food and Drug Administration
- Firm
- Abiomed, Inc.
- Product
- Automated Impella Controller, Packaged. Primary user control interface for the Impella Catheter, catheter-based heart pump.
- Hazard
- Retrospective review of servicing records identified AIC units requiring various hardware, firmware, and/or software update(s): 1. RFID firmware issue that may result in failure of the AIC to recognize the purge cassette during case setup or cassette change. This issue may result in delay in procedure or therapy. 2. Potential for the rotary push button (RPB) sticking and unreliable actuation impacting operator ability to select and confirm menu entries. This can result in loss of control of the console. 3. Potential AIC instability that could result in AIC start-up failure, inability to drive the pump, and the need for console exchange. 4. Software defect related to calculator process crash, leading to Negative Central Venous Pressure (CVP) values present in Pulmonary Artery Pressure (PAP) calculations.
- Units
- 428 units
- Classification
- Class I
- Status
- Ongoing
- Distribution
- Worldwide Distribution: US (nationwide) and OUS (foreign) to countries: Belgium, Brazil, Canada, Colombia, Croatia, Czech Republic, Finland, France, Germany, Hong Kong, India, Ireland, Italy, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Pakistan, Panama, Poland, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom
- Who started it
- Voluntary: Firm initiated
The document
The Food and Drug Administration’s record of the recall is its data, not a document: its record here is the data as read.
What its source changed
Its source has changed nothing since Dorworth News first read it, .
The record as data
The record as data: JSON, 3.1 KB: Abiomed, Inc.: Automated Impella Controller, Packaged. Primary user control interface for the Impella Catheter, catheter-based heart pump
The recall’s record: the product, the hazard, the remedy and the firm in the agency’s own words, its classification and status, and every change the agency made to it.
Source: openFDA (Food and Drug Administration), read . Data provided by the U.S. Food and Drug Administration.