  "id": "dorworth:recall:fda-z-0364-2027",
  "kind": "recall",
  "schemaVersion": 1,
  "schema": "https://dorworth.com/schemas/v1/recall.json",
  "url": "https://dorworth.com/relationships-family/recalls/fda-z-0364-2027",
  "title": "Siemens Healthcare Diagnostics, Inc.: Atellica CH HDL Cholesterol (HDLC); Siemens Material Number (SMN): 11537213",
  "place": [
    {
      "scope": "nation",
      "code": "US",
      "name": "United States"
    }
  ],
  "source": {
    "body": "Food and Drug Administration",
    "url": "https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0364-2027%22",
    "retrievedAt": "2026-11-25T02:45:38.127-05:00"
  },
  "keptCopy": null,
  "published": "2026-11-25T02:45:38.127-05:00",
  "updated": "2026-11-25T02:45:38.127-05:00",
  "changes": {
    "count": 0,
    "url": "https://dorworth.com/changes?record=dorworth%3Arecall%3Afda-z-0364-2027"
  },
  "terms": {
    "credit": "Dorworth News",
    "url": "https://dorworth.com/terms"
  },
  "body": {
    "identifier": "Z-0364-2027",
    "issuedOn": "2026-11-18",
    "issuedBy": "Food and Drug Administration",
    "status": "listed",
    "sourcePage": "https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0364-2027%22",
    "sourceFile": null,
    "copyNote": "The Food and Drug Administration’s record of the recall is its data, not a document: its record here is the data as read.",
    "history": [],
    "agency": "fda",
    "category": "other",
    "firm": "Siemens Healthcare Diagnostics, Inc.",
    "product": "Atellica CH HDL Cholesterol (HDLC); Siemens Material Number (SMN): 11537213;",
    "hazard": "Reagent probe carryover on the Atellica CH and Atellica CI Analyzers may cause falsely elevated Hemoglobin A1c (A1c_E/A1c_H) results due to HDLC carryover (biases of 9 29% in NGSP units), and falsely depressed Direct Bilirubin 2 (DBil_2) results due to TP_2 carryover (biases of 0% to -25%), which may affect patient samples, quality control, and/or calibrator results.",
    "remedy": null,
    "units": "72207 units (9744 US, 62463 OUS)",
    "classification": "Class II",
    "recallStatus": "Ongoing",
    "notes": [
      {
        "text": "Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Burkina Faso, Canada, Chile, China, Colombia, Costa Rica, Croatia, CuraÂ¿ao, CuraÂ¿ao, St Eus, Czech Republic, Democratic Republic Congo, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Maldives, Mexico, Morocco, Nepal, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Turkmenistan, U.A.E., Uganda, United Kingdom , Uruguay, Uzbekistan, Vatikancity, Vietnam.",
        "label": "Distribution"
      },
      {
        "text": "Voluntary: Firm initiated",
        "label": "Who started it"
      }
    ]
  }
}