  "title": "TransMedics, Inc.: OCS Liver Perfusion Set Model/Catalog Number: 3200 Software Version: rev 3 OCS Liver Perfusion Module included in the Liver Perfusion Set …",
  "place": [
    {
      "scope": "nation",
      "code": "US",
      "name": "United States"
    }
  ],
  "source": {
    "body": "Food and Drug Administration",
    "url": "https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0381-2027%22",
    "retrievedAt": "2026-11-25T02:45:38.127-05:00"
  },
  "keptCopy": null,
  "published": "2026-11-25T02:45:38.127-05:00",
  "updated": "2026-11-25T02:45:38.127-05:00",
  "changes": {
    "count": 0,
    "url": "https://dorworth.com/changes?record=dorworth%3Arecall%3Afda-z-0381-2027"
  },
  "terms": {
    "credit": "Dorworth News",
    "url": "https://dorworth.com/terms"
  },
  "body": {
    "identifier": "Z-0381-2027",
    "issuedOn": "2026-11-18",
    "issuedBy": "Food and Drug Administration",
    "status": "listed",
    "sourcePage": "https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0381-2027%22",
    "sourceFile": null,
    "copyNote": "The Food and Drug Administration’s record of the recall is its data, not a document: its record here is the data as read.",
    "history": [],
    "agency": "fda",
    "category": "other",
    "firm": "TransMedics, Inc.",
    "product": "OCS Liver Perfusion Set Model/Catalog Number: 3200 Software Version: rev 3 OCS Liver Perfusion Module included in the Liver Perfusion Set used for organ transport.",
    "hazard": "The Perfusion Module may have an unprogrammed Front End Board (FEB) which will not allow the system to proceed beyond the initial startup process. The OCS system will display an error message saying Unsupported Perfusion Module Present and will prevent further system use.",
    "remedy": null,
    "units": "18",
    "classification": "Class II",
    "recallStatus": "Ongoing",
    "notes": [
      {
        "text": "Worldwide - US Nationwide distribution in the states of CA, FL, IL, MA, MN, NC, NE, NY, TX and the country of Serbia;.",
        "label": "Distribution"
      },
      {
        "text": "Voluntary: Firm initiated",
        "label": "Who started it"
      }
    ]
  }
}