  "id": "dorworth:recall:fda-z-0436-2027",
  "kind": "recall",
  "schemaVersion": 1,
  "schema": "https://dorworth.com/schemas/v1/recall.json",
  "url": "https://dorworth.com/relationships-family/recalls/fda-z-0436-2027",
  "title": "Beyond Air Inc.: NO Calibration Gas Cylinder, 45ppm; Model/Catalog Number: 80006; Software Version: N/A - No firmware/software; Product Description: Nitric …",
  "place": [
    {
      "scope": "nation",
      "code": "US",
      "name": "United States"
    }
  ],
  "source": {
    "body": "Food and Drug Administration",
    "url": "https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0436-2027%22",
    "retrievedAt": "2026-11-25T02:45:38.127-05:00"
  },
  "keptCopy": null,
  "published": "2026-11-25T02:45:38.127-05:00",
  "updated": "2026-11-25T02:45:38.127-05:00",
  "changes": {
    "count": 0,
    "url": "https://dorworth.com/changes?record=dorworth%3Arecall%3Afda-z-0436-2027"
  },
  "terms": {
    "credit": "Dorworth News",
    "url": "https://dorworth.com/terms"
  },
  "body": {
    "identifier": "Z-0436-2027",
    "issuedOn": "2026-11-18",
    "issuedBy": "Food and Drug Administration",
    "status": "listed",
    "sourcePage": "https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-0436-2027%22",
    "sourceFile": null,
    "copyNote": "The Food and Drug Administration’s record of the recall is its data, not a document: its record here is the data as read.",
    "history": [],
    "agency": "fda",
    "category": "other",
    "firm": "Beyond Air Inc.",
    "product": "NO Calibration Gas Cylinder, 45ppm; Model/Catalog Number: 80006; Software Version: N/A - No firmware/software; Product Description: Nitric Oxide Calibration Gas, 45 ppm Intended to be used with LungFit PH device (FDA PMA 200044) during pre verification checks and monthly maintenance to calibration internal NO sensors. Component: No;",
    "hazard": "Some NO calibration tanks may have been contaminated with O2 (from the air).",
    "remedy": null,
    "units": "181",
    "classification": "Class II",
    "recallStatus": "Ongoing",
    "notes": [
      {
        "text": "Worldwide - US Nationwide distribution in the states of California, Tennessee, Arkansas, Nebraska, New York, North Carolina, Delaware, New Jersey, Texas, Indiana, Minnesota, Illinois, Oklahoma, Georgia, Florida, Utah and the countries of Canada, Hong Kong, Australia, Manila, Vietnam, Israel, Taiwan, Panama, Singapore.",
        "label": "Distribution"
      },
      {
        "text": "Voluntary: Firm initiated",
        "label": "Who started it"
      }
    ]
  }
}