Abiomed, Inc.: Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged …
Recall · Food and Drug Administration
In FDA’s enforcement report of Sept. 30, 2026. Class II; Ongoing.
The record
- Recall number
- Z-3251-2026
- Agency
- Food and Drug Administration
- Firm
- Abiomed, Inc.
- Product
- Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US. 5. Impella Optical Controller, Packaged, EU; Product Code: 0042-0010-EU. 6. Impella Optical Controller, Packaged, US; Product Code: 0042-0010-US. 7. Optical, AIC, Impella Connect, Pkgd, US; Product Code: 0042-0040-US.
- Hazard
- Retrospective report of issues remedied in 2019 with firmware and software patches: 1. Defects in optical bench (OPB) firmware V2.19, which affected the accuracy and stability of the pressure signal the AIC system performance depends on. 2. Software defects leading to unreliable or missing placement signal, unresponsive start Impella soft key during disconnect/reconnect, and false Position in Ventricle alarm in certain AICs.
- Units
- 3,322 units
- Classification
- Class II
- Status
- Ongoing
- Distribution
- Worldwide - US Nationwide and the countries of Austria, Belgium, Canada, France, Germany, Hong Kong, Italy, Luxembourg, Poland, Spain, Switzerland, United Kingdom.
- Who started it
- Voluntary: Firm initiated
The document
The Food and Drug Administration’s record of the recall is its data, not a document: its record here is the data as read.
What its source changed
Its source has changed nothing since Dorworth News first read it, .
The record as data
The record as data: JSON, 3.0 KB: Abiomed, Inc.: Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged …
The recall’s record: the product, the hazard, the remedy and the firm in the agency’s own words, its classification and status, and every change the agency made to it.
Source: openFDA (Food and Drug Administration), read . Data provided by the U.S. Food and Drug Administration.