  "id": "dorworth:recall:fda-z-3308-2026",
  "kind": "recall",
  "schemaVersion": 1,
  "schema": "https://dorworth.com/schemas/v1/recall.json",
  "url": "https://dorworth.com/relationships-family/recalls/fda-z-3308-2026",
  "title": "Philips North America: Ingenia 1.5T, 1.5T CX, and 1.5T S; Model number (REF): 1. 781315 (1.5T), 2. 781341 (1.5T), 3. 781396 (1.5T), 4. 782115 (1.5T), 5. 782140 …",
  "place": [
    {
      "scope": "nation",
      "code": "US",
      "name": "United States"
    }
  ],
  "source": {
    "body": "Food and Drug Administration",
    "url": "https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3308-2026%22",
    "retrievedAt": "2026-11-24T20:15:38.497-05:00"
  },
  "keptCopy": null,
  "published": "2026-11-24T20:15:38.497-05:00",
  "updated": "2026-11-24T20:15:38.497-05:00",
  "changes": {
    "count": 0,
    "url": "https://dorworth.com/changes?record=dorworth%3Arecall%3Afda-z-3308-2026"
  },
  "terms": {
    "credit": "Dorworth News",
    "url": "https://dorworth.com/terms"
  },
  "body": {
    "identifier": "Z-3308-2026",
    "issuedOn": "2026-10-07",
    "issuedBy": "Food and Drug Administration",
    "status": "listed",
    "sourcePage": "https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3308-2026%22",
    "sourceFile": null,
    "copyNote": "The Food and Drug Administration’s record of the recall is its data, not a document: its record here is the data as read.",
    "history": [],
    "agency": "fda",
    "category": "other",
    "firm": "Philips North America",
    "product": "Ingenia 1.5T, 1.5T CX, and 1.5T S; Model number (REF): 1. 781315 (1.5T), 2. 781341 (1.5T), 3. 781396 (1.5T), 4. 782115 (1.5T), 5. 782140 (1.5T), 6. 781262 (1.5T CX), 7. 781347 (1.5T S); Software Versions: R11, R12 and R13;",
    "hazard": "When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.",
    "remedy": null,
    "units": "216",
    "classification": "Class II",
    "recallStatus": "Ongoing",
    "notes": [
      {
        "text": "Worldwide distribution - Nationwide and the countries of Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Equator. Guinea, Finland, France, French Guiana, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kyrgyzstan, Latvia, Lebanon, Luxembourg, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Monaco, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Palestine, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Tunisia, T¿rkiye, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Vietnam.",
        "label": "Distribution"
      },
      {
        "text": "Voluntary: Firm initiated",
        "label": "Who started it"
      }
    ]
  }
}