  "id": "dorworth:recall:fda-z-3328-2026",
  "kind": "recall",
  "schemaVersion": 1,
  "schema": "https://dorworth.com/schemas/v1/recall.json",
  "url": "https://dorworth.com/relationships-family/recalls/fda-z-3328-2026",
  "title": "SPECTRUM MEDICAL SRL: Brand Name: Quantum PureFlow Product Name: Quantum PureFlow Standard Heat Exchanger High Flow 3/8 Model/Catalog Number: HX55W-S0 Software …",
  "place": [
    {
      "scope": "nation",
      "code": "US",
      "name": "United States"
    }
  ],
  "source": {
    "body": "Food and Drug Administration",
    "url": "https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3328-2026%22",
    "retrievedAt": "2026-11-24T20:00:38.169-05:00"
  },
  "keptCopy": null,
  "published": "2026-11-24T20:00:38.169-05:00",
  "updated": "2026-11-24T20:00:38.169-05:00",
  "changes": {
    "count": 0,
    "url": "https://dorworth.com/changes?record=dorworth%3Arecall%3Afda-z-3328-2026"
  },
  "terms": {
    "credit": "Dorworth News",
    "url": "https://dorworth.com/terms"
  },
  "body": {
    "identifier": "Z-3328-2026",
    "issuedOn": "2026-10-07",
    "issuedBy": "Food and Drug Administration",
    "status": "listed",
    "sourcePage": "https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3328-2026%22",
    "sourceFile": null,
    "copyNote": "The Food and Drug Administration’s record of the recall is its data, not a document: its record here is the data as read.",
    "history": [],
    "agency": "fda",
    "category": "other",
    "firm": "SPECTRUM MEDICAL SRL",
    "product": "Brand Name: Quantum PureFlow Product Name: Quantum PureFlow Standard Heat Exchanger High Flow 3/8 Model/Catalog Number: HX55W-S0 Software Version: N/A Product Description: The Quantum PureFlow Standard Heat Exchanger is intended to be used with a compatible Heater/Cooler system to heat/cool blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration with a maximum flow rate of 8 l/min. Quantum PureFlow Standard Heat Exchanger is a single use device which allows heating/cooling the blood during extracorporeal circulation. The device is sold sterile in a blister pack in a multiple box of 4 units. Device surfaces in contact with blood are treated with a phosphorylcholine-based biocompatible coating. The device has been designed to be powered by heaters coolers that use water or a glycol-based heat transfer fluid (HTF). Component: N/A",
    "hazard": "Due to blood-to-glycol leak in the heat exchanger.",
    "remedy": null,
    "units": "3232",
    "classification": "Class II",
    "recallStatus": "Ongoing",
    "notes": [
      {
        "text": "US Nationwide distribution in the states of AL, CA, FL, KS MA, MD, MI, MO, NC, NJ, NY, OH, OR, PA, SC, TN, TX, VA, WI.",
        "label": "Distribution"
      },
      {
        "text": "Voluntary: Firm initiated",
        "label": "Who started it"
      }
    ]
  }
}