  "id": "dorworth:recall:fda-z-3330-2026",
  "kind": "recall",
  "schemaVersion": 1,
  "schema": "https://dorworth.com/schemas/v1/recall.json",
  "url": "https://dorworth.com/relationships-family/recalls/fda-z-3330-2026",
  "title": "Tandem Diabetes Care, Inc.: Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version 2.13 …",
  "place": [
    {
      "scope": "nation",
      "code": "US",
      "name": "United States"
    }
  ],
  "source": {
    "body": "Food and Drug Administration",
    "url": "https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3330-2026%22",
    "retrievedAt": "2026-11-24T20:15:38.497-05:00"
  },
  "keptCopy": null,
  "published": "2026-11-24T20:15:38.497-05:00",
  "updated": "2026-11-24T20:15:38.497-05:00",
  "changes": {
    "count": 0,
    "url": "https://dorworth.com/changes?record=dorworth%3Arecall%3Afda-z-3330-2026"
  },
  "terms": {
    "credit": "Dorworth News",
    "url": "https://dorworth.com/terms"
  },
  "body": {
    "identifier": "Z-3330-2026",
    "issuedOn": "2026-10-07",
    "issuedBy": "Food and Drug Administration",
    "status": "listed",
    "sourcePage": "https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-3330-2026%22",
    "sourceFile": null,
    "copyNote": "The Food and Drug Administration’s record of the recall is its data, not a document: its record here is the data as read.",
    "history": [],
    "agency": "fda",
    "category": "other",
    "firm": "Tandem Diabetes Care, Inc.",
    "product": "Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version 2.13 Software Version: iOS software version 2.13 UDI-DI: 00389152470740 Component: No",
    "hazard": "A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous glucose monitoring (CGM) sensors, resulting in the insulin pump being unable to establish a near-field communication (NFC) connection with the CGM and requiring FSL3P users to use their t:slim insulin pumps in manual mode. FSL3P users are unable to access their CGM's glucose values from within the t:slim app and cannot use their pumps in automated control (closed-loop) mode, which could lead to users experiencing hypoglycemia or hyperglycemia.",
    "remedy": null,
    "units": "177",
    "classification": "Class II",
    "recallStatus": "Ongoing",
    "notes": [
      {
        "text": "US Nationwide distribution.",
        "label": "Distribution"
      },
      {
        "text": "Voluntary: Firm initiated",
        "label": "Who started it"
      }
    ]
  }
}