Beckman Coulter Mishima K.K.: Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 / 14987666542965 B96693 / …
Recall · Food and Drug Administration
In FDA’s enforcement report of Oct. 7, 2026. Class II; Ongoing.
The record
- Recall number
- Z-3332-2026
- Agency
- Food and Drug Administration
- Firm
- Beckman Coulter Mishima K.K.
- Product
- Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 / 14987666542965 B96693 / 14987666542972 B12183 / 14987666540671 The AU480 Chemistry Analyzer measure analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material, and other accessories. This system is for in vitro diagnostic use only.
- Hazard
- Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results.
- Units
- 492 units
- Classification
- Class II
- Status
- Ongoing
- Distribution
- Worldwide - U.S. Nationwide distribution in the states of AR, CA, DE, FL, GA, GU, HI, IA, IL, IN, KS, KY, MN, MO, NC, NM, NY, OK, PA, SD, TX, and WI. The countries of Algeria, Australia, Belarus, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Costa Rica, Croatia, Cyprus, Czechia, Egypt, France, Germany, Guatemala, Hungary, India, Ireland, Italy, Japan, Kazakhstan, Korea, Republic of, Latvia, Malaysia, Morocco, Myanmar, Nepal, Nigeria, Pakistan, Poland, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, South Africa, Taiwan, Province of China, Turkey, United Arab Emirates, and Viet Nam.
- Who started it
- Voluntary: Firm initiated
The document
The Food and Drug Administration’s record of the recall is its data, not a document: its record here is the data as read.
What its source changed
Its source has changed nothing since Dorworth News first read it, .
The record as data
The recall’s record: the product, the hazard, the remedy and the firm in the agency’s own words, its classification and status, and every change the agency made to it.
Source: openFDA (Food and Drug Administration), read . Data provided by the U.S. Food and Drug Administration.